Key takeaways
- A deadweight tester is a primary pressure standard. It generates pressure from first principles (F = mg / A) and does not need to be calibrated against another pressure reference; it only needs its mass and piston area to be traceable.
- Digital pressure calibrators (e.g. Fluke 700G series, Druck DPI 620G) achieve 0.025–0.05% FS accuracy. Adequate for calibrating process gauges and transmitters with ±0.5–1% accuracy specifications.
- For accredited laboratory pressure calibration with stated uncertainty below 0.1%, the calibration chain typically uses a deadweight tester as the primary reference. Directly or via a precision transfer standard traceable to it.
- Portable digital pressure calibrators are the standard tool for on-site calibration of pressure gauges, transmitters, and switches in process industries. Fast, battery-powered, and covering gauge, absolute, and differential pressure.
- In Singapore's oil & gas, petrochemical, and pharmaceutical process industries, pressure instrumentation calibration records are reviewed against the required measurement uncertainty , specifying your tolerance determines which calibration method is appropriate.
Pressure calibration fundamentals. Why traceability matters in process industries
Pressure is one of the most widely measured process variables in Singapore's industrial facilities, from pharmaceutical clean rooms monitoring differential pressure across HEPA filters, to oil and gas facilities monitoring line pressure in safety-critical systems, to food processing plants verifying retort and autoclave pressures for sterility assurance. When a pressure gauge or transmitter drifts outside its calibration tolerance, the consequences extend well beyond the measurement itself.
A process gauge reading 10 bar higher than the actual line pressure does not raise an alarm, it gives false confidence. An under-reading gauge may trigger a premature shutdown of a process that should continue, or fail to trigger an alarm at a genuine over-pressure event. In pharmaceutical batch manufacturing, an incorrectly reading autoclave pressure sensor means a batch validation record is wrong. Not just a calibration issue but a GMP compliance failure. In oil and gas, a pressure sensor drift in a safety instrumented system (SIS) loop can mean a protective function does not respond when it should.
The calibration chain for pressure explains why these failures happen and how a rigorous calibration programme prevents them. At the top of the chain sits Singapore's National Metrology Centre (NMC), a division of A*STAR, which maintains the national pressure standards. High-quality deadweight testers whose masses and piston areas are traceable to the international SI system through BIPM comparison programmes. NMC calibrates reference instruments for accredited calibration laboratories, who use those references to calibrate process instruments on the plant floor.
Every step in this chain adds uncertainty. The NMC reference has a small uncertainty; the lab's reference transfer standard inherits that uncertainty and adds a little more; the field instrument calibrated against the lab reference inherits all of those uncertainties combined. This is why the primary standard at the top of the chain (the deadweight tester), matters. Its uncertainty is the floor below which nothing in the chain can go. An unreliable primary standard corrupts every calibration that flows from it. A well-maintained, traceable deadweight tester anchors the entire chain.
How a deadweight tester works. The primary standard for pressure
The deadweight tester (more formally, a piston gauge) is conceptually simple and metrologically elegant. It generates a known pressure from first principles using the equation P = F / A, where F is the gravitational force on a set of calibrated masses (F = mg) and A is the cross-sectional area of the precision-machined piston that carries those masses.
The piston and its matched cylinder are manufactured to extremely tight dimensional tolerances. The piston area is certified and traceable to the national length standard. The weights are individually certified and traceable to the national mass standard. Local gravitational acceleration at the laboratory location is measured or calculated to high precision, in Singapore, g is approximately 9.781 m/s², which differs from the standard 9.80665 m/s² used in some textbooks, and this difference is applied as a correction in every pressure calculation.
In operation, the piston floats in a cylinder filled with hydraulic oil (for higher pressures) or maintained by a gas cushion (for pneumatic deadweight testers at lower pressures). The pressure generated is the pressure at which the piston floats in equilibrium. The weight of the masses divided by the piston area, corrected for fluid head, temperature effects on piston area, and atmospheric pressure. The technician applies weights until the piston floats at the reference height, and that combination of weights defines the calibration point precisely.
Several engineering subtleties make a deadweight tester different from a simple weight-on-a-piston concept. The piston is rotated slowly during measurements (typically at 30–60 rpm), to maintain a thin fluid film between the piston and cylinder, which minimises friction and ensures the piston truly floats rather than sticking. The effective area of the piston is not identical to its geometric area. It depends slightly on the pressure itself (known as piston distortion) and must be characterised as part of the calibration. Temperature affects the piston area through thermal expansion, and corrections are applied from laboratory temperature records. All of these corrections are calculated and applied in accredited laboratory use, contributing to the total uncertainty budget.
The fundamental point (the one that makes the deadweight tester a primary standard), is that the pressure it generates is derived entirely from fundamental SI units. It does not need to be "calibrated" against another pressure device. Its validation is a verification that its masses are correct (comparison against mass standards) and its piston area is correctly characterised (dimensional measurement traceable to the metre). No other pressure instrument it will ever measure is used to confirm its accuracy.
Types of deadweight testers. Oil, pneumatic, and high-pressure variants
Oil deadweight testers are the most common in calibration laboratories and cover the widest pressure range, from approximately 0.1 bar to over 1,000 bar depending on the piston-cylinder combination used. Hydraulic oil serves as the pressure medium, providing incompressibility and piston lubrication. The main disadvantage is contamination risk: if an instrument connected to the oil system is dirty or has incompatible materials, the oil circuit must be cleaned. Oil deadweight testers are bench instruments and cannot be used in the field.
Pneumatic deadweight testers use clean dry nitrogen or air as the pressure medium. They are cleaner, eliminate oil contamination risk, and are preferred for calibrating instruments that cannot tolerate oil, oxygen service gauges, for example. Their practical upper pressure limit is typically around 70 bar; above that, gas compressibility effects and safety concerns make oil systems preferred. Pneumatic deadweight testers are also bench instruments.
High-pressure deadweight testers use specialised piston-cylinder assemblies rated to 10,000 bar or higher and are found in metrology institutes and specialist hydraulic testing laboratories. These cover the pressure ranges needed for hydraulic hose testing, downhole pressure measurement calibration, and ultra-high-pressure process applications.
Key manufacturer models encountered in calibration laboratories include the Fluke/Ruska 2400 series (oil, ranges from 0.01 to 700 bar), the DH Instruments (Ametek) PG7000 series (the reference standard used by many national metrology institutes, available in oil and pneumatic), and the WIKA CPB series (oil and pneumatic, widely used in industrial calibration laboratories). For accredited laboratories, the choice of deadweight tester model is documented in the uncertainty budget and the SAC scope of accreditation.
How digital pressure calibrators work. Precision in a portable package
A digital pressure calibrator is a portable electronic instrument containing a precision pressure transducer (typically piezoresistive (silicon strain gauge) or capacitive), with a temperature-compensated signal chain, an analog-to-digital converter, and a display. The transducer converts applied pressure into a calibrated electrical signal; the signal chain amplifies, temperature-compensates, and linearises it; and the display shows the pressure in engineering units.
The "reference" in a digital pressure calibrator is the transducer inside it, which has been calibrated (typically annually), against a deadweight tester at an accredited laboratory. The calibrator's accuracy specification reflects how closely its readings track the true pressure across its range, temperature range, and over its specified interval between calibrations. This is a secondary or transfer standard: it relies on external calibration rather than deriving its accuracy from first principles.
Digital pressure calibrators come in two broad classes. Measure-only calibrators accept pressure from an external source (the instrument under test's own process, or a hand pump) and display the pressure reading for comparison. Generate-and-measure calibrators include a built-in pump (manual or automated) that generates pressure internally. The Fluke 729 Automatic Pressure Calibrator, for example, includes an automated pressure generation system and a built-in milliamp source/measure for simultaneous loop calibration, enabling a single technician to calibrate a pressure transmitter without a separate pump.
Three instruments represent the spectrum of digital pressure calibrator performance. The Fluke 700G series (pressure modules, 0.05% FS accuracy, 0–700 bar available) is the standard field tool for instrument technicians. The Druck DPI 620G (0.025% FS across multiple pressure ranges, internal reference modules) is the lab-grade portable standard. Used where a higher-accuracy portable reference is needed without resorting to a bench deadweight tester. The WIKA CPH6600 provides similar laboratory-grade performance in a portable housing.
A critical point about "full scale" accuracy specifications deserves emphasis. A calibrator specified at 0.05% FS has an uncertainty of ±0.05% of the full-scale range at every pressure point in that range, not ±0.05% of the reading at each point. On a 0–100 bar module, the uncertainty is ±0.05 bar everywhere: at 100 bar that is ±0.05% of reading; at 1 bar that is ±5% of reading. This asymmetry means that digital pressure calibrators are poorly suited to calibrating low-pressure gauges using a wide-range module. The correct approach is to select a calibrator module whose range closely matches the instrument being calibrated.
Full specification comparison
The table below summarises the key differences between a deadweight tester and a representative digital pressure calibrator for laboratory and field pressure calibration.
| Feature | Deadweight Tester | Digital Pressure Calibrator (Fluke 700G) |
|---|---|---|
| Pressure generation method | Calibrated masses + precision piston (P = mg/A) | Electronic reference transducer |
| Type | Primary standard | Secondary / transfer standard |
| Achievable uncertainty | 0.005–0.02% of reading | 0.025–0.1% full scale |
| Portability | Not portable. Bench instrument | Portable. Battery powered, field-ready |
| Setup time | 15–30 minutes (levelling, fluid selection, priming) | 2–5 minutes |
| Pressure ranges | 0.1 bar to 10,000 bar (depending on model) | 0 to 700 bar (gauge); absolute/differential available |
| Fluid types | Oil (high pressure), water, pneumatic (low pressure) | Pneumatic or hydraulic depending on model |
| On-site use | No | Yes. Standard field tool |
| Pressure media | Hydraulic oil, distilled water | Air/nitrogen (pneumatic), oil (hydraulic) |
| Reference requirements | Mass, piston area. Traceable to NMI | Calibrated against deadweight tester or NMI-traceable reference |
| Calibration of the calibrator | Masses re-verified against NMI; piston area check | Annual calibration against deadweight tester |
| Typical cost | S$15,000 – S$80,000 | S$3,000 – S$8,000 |
| Use case | Accredited lab primary standard, reference gauge calibration | Field calibration, process transmitter verification |
The uncertainty row is the defining difference. A deadweight tester's 0.005–0.02% of reading uncertainty means that at 100 bar, the total uncertainty on the generated pressure is ±0.005–0.02 bar. A digital calibrator at 0.05% FS on a 100 bar range carries ±0.05 bar uncertainty at every point. At most process calibration pressures, both are adequate to calibrate a ±0.5% gauge. But for calibrating a high-accuracy transmitter with a ±0.075% specification, a standard digital calibrator no longer provides the 4:1 test uncertainty ratio that good metrology practice requires, and a deadweight tester or high-grade reference transfer standard is needed.
When to use a deadweight tester. The laboratory calibration case
Deadweight testers are the correct tool in four specific scenarios, and in all of them, the requirement is driven by the need for low uncertainty or by the need to maintain a position in the traceability chain.
As the primary reference in an accredited calibration laboratory. When a calibration laboratory holds SAC-SINGLAS accreditation for pressure, its reference standard for pressure must ultimately be traceable to a primary standard. Almost always a deadweight tester calibrated by or against the NMC. The deadweight tester in the lab is the anchor of its traceability chain. The digital calibrators and reference gauges that field technicians use are calibrated against the lab's deadweight tester at regular intervals. Remove the deadweight tester, and the lab loses its position in the chain.
For calibrating precision pressure instruments with tight tolerance specifications. High-accuracy process transmitters used as reference instruments (for example, a transmitter installed in a pressure calibration rig to verify other transmitters), may have accuracy specifications of 0.04–0.075% of reading or better. Calibrating these requires a reference with an uncertainty better than a quarter of that tolerance, which only a deadweight tester can reliably provide. The same applies to precision test gauges used as portable transfer standards on-site.
For accredited calibration certificates with stated expanded uncertainty below 0.1%. If your quality system, regulatory authority, or customer requires calibration certificates that state expanded uncertainties of ±0.05% of reading or lower, the calibration must use a deadweight-traceable reference. A digital calibrator cannot achieve this level with appropriate test uncertainty ratios.
For calibrating deadweight tester weights and piston areas. The masses in a deadweight tester are sent periodically to a mass calibration laboratory traceable to the NMC; the piston area is characterised through dimensional measurement. This is a different calibration discipline from pressure calibration, but it is what keeps the deadweight tester honest. An accredited lab documents the calibration interval and re-verification evidence for its primary standard.
When to use a digital pressure calibrator. The field calibration case
Digital pressure calibrators are the standard tool for the vast majority of industrial pressure calibration work. The on-site, in-the-plant-at-3am work that keeps Singapore's process industries running. Their advantage is clear: they are portable, battery-powered, quick to set up, and accurate enough for every common process calibration task.
Annual calibration of process pressure gauges and transmitters. Most industrial pressure gauges carry ±1% FS accuracy specifications; smart HART transmitters are typically ±0.065–0.1% of span. A 0.025–0.05% FS digital calibrator provides a 2:1 to 4:1 test uncertainty ratio for these instruments, adequate for standard calibration practice. The field technician connects the calibrator to the instrument under test, applies pressure at 0%, 25%, 50%, 75%, and 100% of range in ascending and descending directions, records as-found and as-left readings, and generates the calibration certificate on-site or back at the lab.
Loop calibration of 4–20mA pressure transmitters. This is where digital pressure calibrators earn their keep in process industries. A transmitter calibration has two variables: the pressure it measures and the 4–20mA signal it outputs. A digital calibrator with an integrated milliamp source and measure function handles both simultaneously. Applying calibrated pressure while sourcing or measuring the loop current, without requiring a separate calibration of the control loop. The Fluke 729 Automatic Pressure Calibrator and the Druck DPI 620G both provide this capability, and it reduces the time and staffing required for transmitter calibration substantially.
Pressure switch verification for safety systems. Safety instrumented system (SIS) pressure switches have a defined set point at which they must activate. Verifying set point accuracy requires applying increasing pressure until the switch activates and recording the trip pressure. A portable digital calibrator handles this quickly and accurately for common process switch specifications.
New instrument commissioning and verification. Before installing a new pressure transmitter in a process loop, confirming that it reads correctly across its range (and is configured correctly in the DCS or SCADA system), is standard commissioning practice. A portable digital calibrator performs this loop check on-site, saving a return trip to the calibration laboratory.
Pressure gauge and transmitter calibration. Traceable to NMC, stated uncertainty on every certificate
Unitest's pressure calibration is traceable to deadweight tester references calibrated against NMC Singapore. Accredited certificates state expanded uncertainty. Ready for HSA GMP, HACCP, and ISO 9001 audits.
Pressure calibration in Singapore's regulated industries
Singapore's regulated industries impose calibration requirements on pressure instruments that often go beyond what a company's internal quality system requires, and the calibration certificate must satisfy not just the internal auditor but an external regulatory inspector who may scrutinise the traceability statement line by line.
Pharmaceutical manufacturing (HSA GMP). The Health Sciences Authority's GMP guidelines, aligned with PIC/S and ICH Q10, require critical process instruments to be calibrated with traceable reference standards and documented uncertainties. For pressure instruments in clean rooms (differential pressure monitoring), sterilisation systems (autoclaves, steam-in-place), and formulation equipment (mixing vessel pressure, filling line pressure), calibration certificates must demonstrate traceability. HSA inspectors performing Annex 15 qualification reviews and periodic GMP audits look for uncertainty statements on pressure calibration certificates. Missing uncertainty data is a common GMP observation in Singapore pharmaceutical facilities.
Food manufacturing (HACCP and SFA). Retort processing for commercially sterile food products requires that the process pressure (one of the lethality parameters alongside temperature and time), is monitored with calibrated, traceable instruments. HACCP plans list pressure gauges on retorts and autoclaves as critical control point instruments, and auditors under ISO 22000, FSSC 22000, and SQF require calibration records. The Singapore Food Agency (SFA) also reviews calibration documentation during facility audits for licensed food manufacturers.
Oil and gas and petrochemical (MOM, WSHO). The Workplace Safety and Health Act and associated regulations in Singapore's process industries require that safety instrumented systems (including pressure transmitters in SIS loops), are maintained and tested. Pressure instruments in Category 1 SIS loops require calibration evidence as part of functional safety management. The calibration certificate must be traceable, state the range and uncertainty, and be retained in the instrument's maintenance record. For SIL-rated instruments, calibration tolerance is typically determined by the SIL assessment. The required probability of failure on demand (PFD) informs the required measurement accuracy, which in turn determines the calibration method.
What auditors specifically look for. An experienced regulatory auditor reviewing a pressure calibration certificate checks: the instrument identification and serial number (does it match the instrument installed?), the calibration date and due date (is it currently in-date?), the pressure range covered (does the calibration cover the instrument's operating range?), the reference standard used and its traceability (is it traceable to NMC?), and the stated expanded uncertainty (is the uncertainty stated? Is it adequate for the instrument's tolerance?). A certificate that states only "pass/fail" or provides readings without uncertainty frequently triggers an observation. Not because the calibration was performed poorly, but because the certificate cannot prove it was performed adequately.
Common mistakes in pressure calibration
Five calibration mistakes appear repeatedly in Singapore process industry calibration programmes, and understanding them prevents the most common sources of calibration non-conformance.
(a) Using a wide-range calibrator for low-pressure instruments. A 0–700 bar digital calibrator module used to calibrate a 0–1 bar gauge has an uncertainty of ±0.35 bar , 35% of the gauge's full range. This is not a useful calibration. Select a calibrator module or range that matches the instrument: a 0–1 bar module for a 1 bar gauge, a 0–10 bar module for a 10 bar gauge. The accuracy specification is always referenced to the full scale of the module in use.
(b) Ignoring head pressure corrections. When the deadweight tester in the calibration laboratory is at a different elevation from the connection point of the instrument under test (or when a long vertical tube filled with oil or water connects the calibrator to the instrument), the hydrostatic head of the fluid column adds to or subtracts from the applied pressure. At 1 metre of water column, the head pressure is approximately 0.1 bar. A significant error for a ±0.5% gauge on a 10 bar range, and a critical error for a low-pressure gauge on a 0–1 bar range. Accredited calibration laboratories account for head pressure corrections in their uncertainty budgets; field calibrations often do not.
(c) Calibrating at a single point only. A pressure gauge can be accurate at zero and at full scale but significantly non-linear in the mid-range. Particularly Bourdon tube gauges that have aged or been over-pressured. A single-point calibration at the midpoint misses the span and zero errors; a single-point at full scale misses the mid-range non-linearity. Five-point calibration (0%, 25%, 50%, 75%, 100% ascending and descending) is standard practice for industrial gauges and is required for accredited calibration certificates that cover the full range.
(d) Not accounting for the gauge medium. A pressure gauge originally calibrated with hydraulic oil and then used to measure steam pressure experiences a head pressure difference because steam (at process temperature) has a very different density from the oil column in the calibration setup. Similarly, a gauge calibrated with nitrogen and then used to measure hydraulic oil pressure must account for the different media. Medium mismatch errors are common in facilities where instruments are recalibrated in a central lab but return to a different service environment.
(e) Using a digital calibrator outside its rated temperature range. A digital pressure calibrator's accuracy specification is stated at a reference temperature, typically 18–28°C for most instruments. Singapore's ambient temperatures can reach 35–38°C in non-air-conditioned plant areas, and direct sunlight raises instrument temperatures higher still. Most digital calibrators have temperature compensation algorithms that reduce (but do not eliminate), the temperature effect. Calibrating in a 38°C environment with a calibrator rated at 0.05% FS at 23°C may produce an effective uncertainty of 0.08–0.1% FS. Sufficient to change whether a marginally non-conforming instrument appears to pass. When calibrating in hot environments, allow the calibrator to equilibrate, shield it from direct sun, and check whether temperature effect specifications cover your operating conditions.
Unitest's pressure calibration capability and what certificates cover
Unitest Instruments holds SAC-SINGLAS accreditation (ISO/IEC 17025, no. LA-2023-0845-C) covering pressure calibration. The accredited scope includes pressure gauges (Bourdon tube, capsule, and diaphragm types), digital pressure indicators, pressure transmitters (including 4–20mA and HART), and differential pressure transmitters across a range of gauge and absolute pressure ranges. The full scope document, including the specific parameters, ranges, and stated expanded uncertainties, is available on the SAC website and as a downloadable PDF from Unitest's accreditation page.
Every Unitest accredited pressure calibration certificate states: the instrument identification and serial number, the calibration date, the due date (based on the agreed calibration interval), the reference standard used and its traceability, the measured values at each calibration point in both ascending and descending directions, the as-found and as-left data (confirming whether the instrument was adjusted during calibration), and the expanded measurement uncertainty for the applied pressure. Calculated from the combined uncertainties of the reference standard, the measurement process, and the resolution of both instruments.
For most industrial gauge and transmitter calibration in Singapore, Unitest's laboratory calibration against deadweight-traceable references provides appropriate uncertainty at competitive cost and turnaround time. Typically 2–5 working days for standard instruments. For facilities requiring on-site calibration of process instruments that cannot be removed from service, Unitest's field calibration service uses SAC-traceable digital reference calibrators, with certificates that clearly state the reference standard's traceability and the achievable on-site uncertainty.
If your process requires stated expanded uncertainty below 0.02% FS for a specific pressure range (for example, calibrating a precision differential pressure transmitter used as a flow measurement reference), confirm the specific capability for that range with Unitest before submission. The achievable uncertainty depends on the pressure range, the instrument type, and the reference standard appropriate for that range. For most process instrumentation in Singapore's manufacturing and process industries, the standard laboratory service provides appropriate uncertainty with full documentation for audit purposes.
Frequently asked questions
A primary pressure standard is a device that generates a known pressure from first principles (derived directly from fundamental SI units (mass, length, time)), without reference to any other pressure instrument. A deadweight tester is a primary standard because the pressure it generates is calculated from the mass of the weights (traceable to the kilogram), the piston area (traceable to the metre), and local gravitational acceleration. This means its accuracy does not depend on any other pressure device. Every pressure calibration chain ultimately traces back to a primary standard like a deadweight tester at a national metrology institute such as Singapore's NMC. Without a reliable primary standard at the top of the chain, every calibration that flows from it carries an unquantified, unverified error.
Yes, substantially. A well-maintained deadweight tester achieves expanded uncertainty of 0.005–0.02% of reading. A high-quality portable digital pressure calibrator such as the Druck DPI 620G achieves 0.025% of full scale, and a field-grade instrument like the Fluke 700G achieves 0.05% of full scale. For a 0–100 bar digital calibrator with 0.05% FS accuracy, the uncertainty is ±0.05 bar at every point, at 1 bar, that represents ±5% of reading. The digital calibrator's own accuracy depends on its internal reference transducer, which was itself calibrated against a deadweight tester. A digital calibrator's uncertainty is always larger than the deadweight tester used to calibrate it.
For calibrating standard process pressure gauges and transmitters with ±0.5–1% accuracy specifications, a 0.025–0.05% FS digital pressure calibrator (Fluke 700G, Druck DPI 620G) provides adequate reference uncertainty. Typically a 4:1 test uncertainty ratio or better. For higher-accuracy process transmitters (0.075% FS or better), or for accredited calibration with stated uncertainty below 0.1%, the laboratory should use a deadweight-traceable reference. To determine whether your calibrator is adequate for a specific instrument, calculate the test uncertainty ratio (TUR): the calibrator's uncertainty should be at least four times better than the gauge's tolerance. If you are unsure, Unitest can advise on the appropriate calibration approach for your specific instruments.
NMC is the National Metrology Centre, Singapore's national metrology institute operated by A*STAR. "Traceable to NMC" for pressure means there is an unbroken, documented chain of calibrations connecting your instrument's certificate back to the pressure standards maintained by the NMC. Typically NMC's own deadweight testers, which are calibrated against BIPM-linked international comparisons. Each link in the chain adds a small, quantified uncertainty. An accredited lab like Unitest calibrates its reference pressure instruments directly against NMC, then uses those references to calibrate customer instruments. The traceability statement on the certificate confirms this chain exists, is documented, and that every uncertainty contribution has been accounted for in the stated expanded uncertainty.
There is no single universal interval. The correct interval depends on the instrument's stability history, the criticality of the measurement, and the required confidence level. A common starting point for industrial pressure gauges in process industries is 12 months. High-wear applications (steam service, high-cycle switching, corrosive media) or safety-critical instruments (SIS loop pressure sensors) may require 6-month or quarterly intervals. Instruments with a documented history of staying well within tolerance can often be extended to 18 or 24 months with written justification. ISO 9001:2015 clause 7.1.5 requires calibration intervals to be set and reviewed based on evidence. Not simply following a manufacturer recommendation. Review as-found data at each calibration to determine whether the interval is appropriate for your specific instruments and conditions.
Yes. Unitest's SAC-SINGLAS accredited pressure calibration scope covers pressure gauges (Bourdon tube, digital, capsule, and diaphragm types), pressure transmitters including 4–20mA two-wire and smart (HART) transmitters, differential pressure transmitters, and pressure switches. Calibration of transmitters includes applying calibrated pressure at multiple points across the measurement range and recording both the pressure error and the output signal error. Certificates state the expanded uncertainty for both the applied pressure and the output signal reading. Confirm your specific instrument type, pressure range, and output type with Unitest before submission to ensure it falls within the accredited scope for your certificate requirements.
Loop calibration combines pressure calibration of the transmitter with simultaneous verification of its 4–20mA output signal, in a single pass. A digital pressure calibrator with a built-in milliamp source and measure function (such as the Fluke 729 Automatic Pressure Calibrator or the Druck DPI 620G) applies known pressure while simultaneously measuring or sourcing the 4–20mA loop current. This eliminates the need for a separate milliamp meter and allows both the pressure accuracy and the output signal accuracy to be verified together. Loop calibration is the standard method for instrument technicians calibrating pressure transmitters in pharmaceutical, oil and gas, and food processing plants across Singapore. The calibration certificate records the as-found and as-left condition for both applied pressure and mA output at each calibration point.
Pressure gauge and transmitter calibration. Deadweight-traceable, SAC-SINGLAS accredited
Unitest calibrates pressure gauges, transmitters, and switches against deadweight-traceable references. Certificates state expanded uncertainty in Pascal, bar, and psi, ready for process industry audits.
Verifiable at sac.gov.sg · Acc. No. LA-2023-0845-C