SAC-SINGLAS Accredited ISO/IEC 17025 Acc. No.LA-2023-0845-C Traceable to Singapore's NMC View accreditation
Complete Guide

The Complete Guide to Instrument Calibration in Singapore (2026)

What calibration is, what it is not, how SAC-SINGLAS accreditation works, what ISO 9001 actually requires, how to set calibration intervals, what must appear on a certificate, and what to do when an instrument fails, everything in one place.

Unitest Editorial15 min readWritten by an ISO/IEC 17025 accredited lab
Electrical instrument calibration at a SAC-SINGLAS accredited laboratory in Singapore
The short answer Calibration is the process of comparing a measuring instrument's output against a traceable reference standard to determine and record any deviation from the expected value. In Singapore, calibration is governed by the Singapore Accreditation Council (SAC) through its SINGLAS accreditation scheme for calibration laboratories, aligned with ISO/IEC 17025. This guide covers everything from the basic definition of calibration through to ISO 9001 compliance, SAC-SINGLAS accreditation, calibration intervals, out-of-tolerance procedures, calibration certificates, and how to choose a calibration lab in Singapore.

Key takeaways

  • Calibration does not adjust an instrument. It determines how much the instrument deviates from a traceable reference; adjustment is a separate action taken after finding a deviation, and it is optional.
  • SAC-SINGLAS accreditation (ISO/IEC 17025) is Singapore's nationally recognised standard for calibration laboratories. A certificate from an accredited lab is traceable to SI units through NMC Singapore; a certificate from an unaccredited source is not.
  • ISO 9001 clause 7.1.5.2 requires calibration of measuring equipment used in quality-affecting activities with traceable certificates, but it does not specify intervals; the interval is the organisation's responsibility to determine based on risk.
  • An out-of-tolerance instrument found during calibration requires a documented impact assessment covering the period since the last calibration. Results produced during that period must be reviewed for potential defects.
  • The calibration certificate must state the measurement result, reference standard, expanded uncertainty, and pass/fail against specification. A "CALIBRATED" sticker without a certificate is not compliant with ISO 9001, IATF 16949, or ISO 13485.

Calibration at a glance. Reference table

Before the detail, here is the one-page summary that covers the most common questions quality managers and engineers ask about calibration in Singapore.

Topic The short answer
What is calibration Comparison of an instrument to a traceable reference to determine deviation
What calibration is NOT Not adjustment, repair, or certification that the instrument is "good"
Accreditation in Singapore SAC-SINGLAS under ISO/IEC 17025, traceable to NMC Singapore
ISO 9001 requirement Clause 7.1.5.2. Calibrate with traceable certificates, retain records
IATF 16949 requirement Clause 7.1.5.3.1. Annual calibration plan, GRR studies, certified labs
Calibration interval Determined by the user; typically 12 months for most instruments
Out-of-tolerance procedure Quarantine → impact assessment → root cause → corrective action
Calibration certificate Must state result, uncertainty, standard, date, lab accreditation number
When to send for calibration Before first use, at defined intervals, after impact/damage/anomaly
Calibration vs verification Calibration = measure deviation; verification = check the result is within spec

1. What calibration is, and what it is not

Calibration is a documented comparison: a measuring instrument is measured against a traceable reference under defined conditions, and the deviation is recorded. That is all calibration does. It documents the truth about the instrument's current state.

Calibration does not: adjust the instrument, certify it as "good," guarantee future accuracy, or repair it. Adjustment is a separate action. An adjusted instrument is then re-calibrated to document the new state after the adjustment. The calibration record shows where the instrument was before adjustment and where it is after.

Understanding this distinction prevents the most common operational mistake: sending an instrument for "calibration" and expecting it to come back precisely set. It comes back with a record of where it was, and if adjustment was requested and performed, a further record of where it is now. If the instrument is outside its specification after calibration, it needs adjustment. Calibration alone will not fix it.

The other common confusion is between calibration and verification. Calibration measures the deviation. Verification is the act of comparing that deviation against a tolerance to decide whether the instrument is fit for its intended use. You can calibrate an instrument and find it deviates by 0.3°C (that is calibration. Deciding whether 0.3°C deviation is acceptable for your 1°C-tolerance cold room process), that is verification.

2. Why calibration matters in Singapore

Singapore's manufacturing and services sector uses measurement in every production process: temperature in pharmaceutical cold rooms, pressure in chemical reactors, current in power distribution, dimensional measurements in precision machining, weighing in food production. A measurement error propagates through the entire downstream process.

Consider a temperature sensor that reads 2°C low. A cold room that displays 4°C is actually at 6°C. Pharmaceutical product stored there may be outside its 2–8°C specification while appearing compliant on the monitoring system. The error is invisible until calibration reveals it. Or until a product failure, a recall, or a regulatory inspection reveals it in a much more costly way.

Singapore's quality infrastructure (ISO 9001, IATF 16949, ISO 13485, GMP, HACCP), relies on calibrated instruments to assure product quality at every stage. Without calibration, measurement data cannot be trusted, and quality assurance becomes a documentation exercise rather than a technical control. For Singapore manufacturers exporting to regulated markets, calibration evidence that satisfies overseas customers and regulators starts with a traceable chain to Singapore's National Metrology Centre (NMC) through SAC-SINGLAS accredited laboratories.

3. The calibration traceability chain

Traceable calibration means the measurement result is connected to SI units through an unbroken chain of calibrations, each with a known and documented uncertainty.

The chain works as follows: SI units are defined by fundamental physics. The metre by the speed of light, the kilogram by Planck's constant, the kelvin by Boltzmann's constant, the ampere by the elementary charge. These definitions are universal and reproducible anywhere in the world. Singapore's National Metrology Centre (NMC), a division of A*STAR, maintains primary standards that realise these SI units in practical measurement instruments. NMC's standards are compared internationally through key comparisons with other national metrology institutes to confirm their accuracy.

From NMC, the chain continues to SAC-SINGLAS accredited calibration laboratories, which maintain secondary and working reference standards calibrated directly against NMC. These are the labs that calibrate your instruments. Each link in the chain adds a known uncertainty contribution, documented in the uncertainty budget on the calibration certificate. The final link (your instrument calibrated against the lab's working standard), has a stated uncertainty that reflects the combined contributions of every link above it.

A calibration certificate from a non-accredited source may claim traceability, but the claim cannot be independently verified. The accreditation assessment is precisely the mechanism that confirms traceability is real (not asserted), at every link in the chain. This is why SAC-SINGLAS accreditation, not self-declaration, is the meaningful standard in Singapore's quality ecosystem.

4. SAC-SINGLAS accreditation. What it means

SAC-SINGLAS (Singapore Laboratory Accreditation Scheme) is operated by the Singapore Accreditation Council (SAC) under A*STAR. The standard for calibration laboratory competence is ISO/IEC 17025:2017. Accreditation to this standard means an independent assessor has verified five areas of technical competence.

Technical competence of personnel

Every technician performing accredited calibrations must demonstrate competence for the specific parameters they work with. Through training records, internal authorisation, and technical assessment. The accreditation assessment verifies that competence is documented and maintained, not assumed.

Suitability of equipment and facilities

Reference instruments must be appropriate for the parameter and range, maintained to the required performance level, and themselves calibrated on schedule. Environmental conditions during calibration (temperature, humidity, vibration), must be monitored and recorded. SAC assessors inspect equipment logs, environmental monitoring records, and the calibration facilities directly.

Valid calibration methods

The lab must use technically validated measurement methods appropriate to each parameter. Methods outside the accredited scope cannot be legitimately covered by the accreditation reference on the certificate. Accreditation is scope-specific, and the scope is publicly published on sac.gov.sg.

Measurement uncertainty quantification

Every accredited certificate must state the measurement uncertainty for each result. SAC assessors technically review whether uncertainty budgets are correctly identified, calculated, and reported. This is the most technically demanding aspect of the assessment and the one most frequently missed by non-accredited providers.

Documented quality management system

The lab must maintain a quality system covering all calibration activities. Procedures, records, corrective actions, and internal audits. This is not the same as ISO 9001 certification: the quality system here is calibration-specific, not general.

A SAC-SINGLAS accredited laboratory is assessed by SAC against all five areas on-site, not merely through document submission. Assessments are conducted at initial accreditation, through annual surveillance visits, and full reassessments on a cycle. SAC is a signatory to APLAC (Asia Pacific Laboratory Accreditation Cooperation) and ILAC (International Laboratory Accreditation Cooperation) mutual recognition arrangements, meaning certificates from SAC-SINGLAS labs are internationally recognised. Accepted in Japan, EU, USA, Australia, and over 100 other economies without re-calibration.

5. ISO 9001 calibration requirements, clause 7.1.5

ISO 9001:2015 clause 7.1.5 (Monitoring and measuring resources) is the primary calibration requirement for most Singapore quality management systems. It operates at two levels.

Clause 7.1.5.1 (General) requires that when measurement is necessary to demonstrate product or service conformity, the organisation must determine and provide suitable monitoring and measuring resources. Those resources must be maintained in a manner that ensures fitness for purpose.

Clause 7.1.5.2 (Measuring equipment) requires specifically:

  • Calibration or verification against traceable measurement standards at specified intervals, or before use
  • Identification to determine calibration status
  • Safeguarding from adjustment, damage, or deterioration that would invalidate calibration status
  • Retention of documented information as evidence of fitness for purpose

The 2015 revision added explicit language about measurement uncertainty. The standard now requires that calibration produce results with stated measurement uncertainties, assessed against the measurement requirement. An ISO 9001 auditor trained to the current standard will ask to see uncertainty on calibration certificates. A certificate that states only pass/fail results without uncertainty gives the auditor grounds to raise a non-conformance.

One critical point that is frequently misunderstood: ISO 9001 does not specify calibration intervals. The standard says "at specified intervals or before use". The interval is the organisation's responsibility to determine. The auditor will check that intervals exist and are documented; they will not tell you that 12 months is right. Risk-based thinking (clause 6.1) should inform the interval decision.

6. Other Singapore quality standards requiring calibration

Beyond ISO 9001, several sector-specific standards impose calibration requirements with additional rigour.

IATF 16949 (automotive)

Clause 7.1.5.3.1 requires an annual calibration and measurement system analysis plan covering all measuring and monitoring equipment used to verify product conformity. Calibration must be performed by laboratories with demonstrated calibration competence, for external laboratories, this means SAC-SINGLAS accreditation or equivalent national accreditation. Clause 7.1.5.3.1 also requires Gauge Repeatability and Reproducibility (GRR) studies for all measurement systems used in production and product conformity verification.

ISO 13485 (medical devices)

Clause 7.6 requires calibration at defined intervals, using equipment traceable to national or international measurement standards. The interval must be documented, and out-of-tolerance procedures must be defined and followed. Software used for monitoring and measurement must be validated before use and re-validated when modified. Records of calibration results must be retained for at least the lifetime of the medical device.

AS9100 (aerospace)

Clause 7.1.5 adds requirements for measurement system analysis, tool control, and first article inspection to the ISO 9001 baseline. Calibration evidence must demonstrate that measurement systems are capable of meeting the accuracy required for their intended use, which typically demands a tighter test uncertainty ratio (TUR) than general industry.

GMP pharmaceutical

Each regulatory body (HSA/PIC/S, FDA, EMA), has specific calibration requirements for instruments used in manufacturing and quality control. Critical instruments (balances, temperature sensors, pressure gauges, pH meters) must be calibrated with traceable references, with defined calibration intervals, and with documented evidence retained in the instrument qualification file. Annex 15 (qualification and validation) requires calibration of all instruments used in installation qualification (IQ), operational qualification (OQ), and performance qualification (PQ) activities.

7. Calibration intervals. How to set them

The calibration interval is the time between calibrations. The interval must be set by the instrument user, not the calibration laboratory. Setting it correctly is a risk-based judgement, not a default.

Several factors determine the appropriate interval for a given instrument:

  • Instrument stability and drift rate. How fast does this type of instrument drift? This is known from calibration history (multiple calibration records for the same instrument over time) and from manufacturer specifications. An instrument with a consistent 0.05°C per year drift history supports a longer interval than one that drifted 0.3°C in six months.
  • Severity of the application. A temperature sensor controlling a pharmaceutical cold room at 2–8°C is critical; a temperature indicator in a staff kitchen is not. The consequence of an undetected out-of-tolerance condition should drive interval decisions.
  • Manufacturer's recommendation. A useful starting point, but not a substitute for data from actual calibration history. Manufacturers specify intervals for typical use; your application may be more or less demanding.
  • Cost of calibration vs. cost of error. Shortening an interval costs money; extending it risks using an out-of-tolerance instrument. The analysis should be explicit, not intuitive.

Common intervals used in Singapore practice: 12 months for most industrial instruments; 6 months for instruments in critical safety service or with demonstrated drift history; 3 months for precision pharmaceutical, semiconductor, or aerospace environments; 24 months for demonstrated-stable instruments where calibration history provides evidence of slow and consistent drift.

ISO 10012:2003 (Measurement management systems) provides a formal framework for setting and adjusting intervals based on calibration results. The approach: review the calibration history after each cycle; if the instrument consistently passes with large margin, consider extending the interval; if it shows trends or approaches out-of-tolerance, reduce the interval or investigate the cause.

SAC-SINGLAS Calibration. All Parameters

One lab for all your calibration needs. SAC-SINGLAS accredited, NMC-traceable

Unitest calibrates electrical, temperature, pressure, dimensional, and flow instruments against NMC-traceable references. SAC-SINGLAS accredited laboratory serving Singapore's manufacturing, pharmaceutical, and process industries.

8. The calibration certificate. What must be on it

A compliant calibration certificate under ISO/IEC 17025:2017 is a specific document with defined required content. Many certificates in circulation are incomplete, and an incomplete certificate carries the same audit risk as no certificate at all, because the missing elements are precisely the ones auditors check.

A complete calibration certificate must include:

  • Laboratory identification. Name, address, and SAC-SINGLAS accreditation number
  • Certificate identification. Unique certificate number and total number of pages
  • Dates. Date of calibration (when the measurement was performed) and date of issue (when the certificate was produced); these may differ
  • Instrument under test. Name, manufacturer, model, serial number, and any other identifier necessary to unambiguously identify the instrument
  • Environmental conditions. Temperature, humidity, and pressure recorded during calibration, where these affect the result
  • Reference standards. Identification of the measurement standards and reference instruments used, with their own calibration certificate numbers and traceability
  • Calibration results. The actual measurement readings at each test point, expressed in SI units or units derived from SI
  • Measurement uncertainty. The expanded uncertainty for each result, with the coverage factor k and confidence level (typically k=2, 95%)
  • Pass/fail statement. If the client has provided a specification, the certificate should state whether the instrument meets that specification, taking uncertainty into account
  • Authorised signature. Signature of the technically responsible person, or electronic equivalent
  • Statement of accreditation, whether the results are covered by the lab's accreditation scope, and for which parameters

A "CALIBRATED" sticker attached to an instrument is an administrative convenience, not a quality record. If the sticker is the only documentation available when an auditor asks for calibration evidence, the result is a non-conformance. The certificate is the record; the sticker only points to the certificate.

9. Out-of-tolerance procedure. What to do when calibration fails

An out-of-tolerance (OOT) finding during calibration is a quality event that requires a structured response. This is one of the most frequently mishandled situations in calibration management. The instrument is often simply sent for repair without the mandatory impact assessment.

Step 1: Quarantine immediately

Label the instrument "OUT OF SERVICE. DO NOT USE" and physically remove it from the work area. If it is connected to a process, remove or bypass it. No further measurements should be made with the instrument until the OOT situation is resolved.

Step 2: Determine the affected period

The affected period is the time since the last successful calibration. Typically the date of the previous calibration certificate. All measurements made with this instrument during that period are potentially affected and must be reviewed.

Step 3: Impact assessment

Review all measurement records from the affected period. For each record, assess: was this measurement used in a quality-affecting decision? What is the maximum possible measurement error given the OOT deviation? What is the consequence of that error on the product or test result? Document the assessment and its conclusions. If product may have been adversely affected, follow your non-conforming product procedure. This may include customer notification, product quarantine, or recall, depending on the severity.

Step 4: Root cause analysis

Determine why the instrument drifted out of tolerance. Common causes include physical shock or impact, exposure to conditions outside the instrument's rated environment (temperature, humidity, vibration), degraded components, inherent drift exceeding the calibration interval, and incorrect handling or storage. The root cause determines the corrective action and whether the calibration interval needs adjustment.

Step 5: Corrective action

Replace, repair, or adjust the instrument as appropriate, then recalibrate to confirm it is now within tolerance. Implement preventive measures to address the root cause. Protective cases, environmental controls, handling procedures, or shorter calibration intervals. Update the calibration register with the new calibration due date.

Step 6: Documentation

The entire OOT response (the impact assessment, root cause analysis, corrective action, and outcome), constitutes a mandatory quality record. ISO 9001 clause 8.7 (non-conforming outputs) and clause 10.2 (nonconformity and corrective action) both apply. Retain all records. The OOT event and its resolution will be reviewed during your next ISO 9001 surveillance audit.

10. Choosing a calibration lab in Singapore

Not all calibration providers in Singapore offer the same level of evidence, traceability, or technical competence. The selection decision has direct consequences for your quality system's audit readiness. Here are the questions to ask before committing.

Is the lab SAC-SINGLAS accredited?

This is the threshold question. Ask for the accreditation number and verify it on sac.gov.sg. The verification is free and takes under two minutes. Check that the accreditation is currently active. Not lapsed, suspended, or pending renewal. Any lab that cannot provide a verifiable accreditation number should be treated as non-accredited.

Does their scope cover your instrument?

Accreditation is scope-specific. A lab accredited for electrical calibration may not be accredited for pressure or temperature. The scope document, available on sac.gov.sg, lists the specific parameters, measurement ranges, and measurement uncertainties the lab is accredited to calibrate. Confirm that your instrument's parameter and the range you need calibrated both fall within the accredited scope. A certificate covering a parameter outside the scope is not an accredited certificate, even if the accreditation number appears on the document.

What reference standards do they use?

The lab's working reference standards should be traceable to NMC Singapore, with current calibration certificates. Ask when their reference standards were last calibrated and by whom. A lab whose references have lapsed calibration cannot provide traceable results, regardless of their accreditation status.

What is their measurement uncertainty?

The lab's measurement uncertainty for your parameter should be at least 3–4 times better than your instrument's specification, the test uncertainty ratio (TUR). A 4:1 TUR is the minimum for most industrial applications; tighter processes may require 10:1. If the lab cannot state its measurement uncertainty for your parameter before you send the instrument, that is a warning sign.

Can they accommodate your specific instrument?

Some instruments require special fixturing, environmental conditioning, specific software interfaces, or unusual procedures. Confirm before sending that the lab has the capability and experience to calibrate your model, not just your parameter type in general.

What is the turnaround time?

Standard turnaround in Singapore is typically 5–10 working days for most parameters. Express service (24–48 hours) is available from some labs, including Unitest, for standard electrical and temperature parameters. If your production schedule cannot accommodate standard turnaround, confirm express availability and pricing before committing.

11. Common calibration mistakes in Singapore

Quality managers and engineers across Singapore's manufacturing and service sectors encounter the same calibration errors repeatedly. Recognising them prevents the non-conformances, audit findings, and production consequences that follow.

Calibrating only the sensor, not the complete measurement system

A temperature measurement system consists of a sensor (thermocouple, PT100) and a transmitter or indicator that converts the sensor signal to a displayed reading. Both contribute to the total measurement error. Calibrating only the sensor tells you nothing about the accuracy of the displayed value. The transmitter may have its own drift. For temperature systems used in quality-affecting applications, calibrate the complete loop: sensor plus transmitter, together, under conditions representative of actual use.

Using stickers as the only calibration record

Calibration due-date stickers serve one purpose: giving a technician a quick visual check that an instrument is within its calibration window. They are not quality records. When an auditor asks for calibration evidence (or when an instrument goes OOT and you need the impact assessment), the sticker provides nothing. Maintain a written or electronic calibration register as the primary record, with certificate numbers, dates, results, and due dates for every instrument.

Accepting certificates without measurement uncertainty

A certificate that states only that the instrument "passed" or gives raw readings without uncertainty is an incomplete certificate under ISO/IEC 17025:2017. Uncertainty of ±0.05°C means something fundamentally different from ±0.5°C. The same reading can be either fit or unfit for purpose depending on which applies. Always check that the certificate states the expanded uncertainty, coverage factor k, and confidence level for every measurement result. If it does not, the certificate does not satisfy ISO 9001:2015 clause 7.1.5 as written.

Calibrating over a range narrower than the instrument's working range

If an instrument is used to measure 0–100°C, calibrating it only at room temperature (20–25°C) tells you nothing about its accuracy at the extremes of its working range. Calibration must cover the full range. Or at minimum the range actually used in quality-affecting measurements. Document the calibration range and confirm it covers the instrument's actual application.

Allowing calibration to expire without renewal

Once a calibration due date passes, the instrument's measurement status is unknown. Using it for quality-affecting measurements from that point is the same as using an uncalibrated instrument. It creates a potential OOT situation with a retroactive affected period. Set up automated reminders in your calibration register at least 4 weeks before the due date to allow time for scheduling, transport, calibration, and return before the due date arrives.

12. Summary. The practical calibration checklist for Singapore

Calibration in Singapore operates within a well-structured national system: NMC provides the primary standards; SAC-SINGLAS accredits laboratories to ISO/IEC 17025; your quality system requires you to use calibrated instruments with traceable certificates. The practical implementation comes down to six steps.

  1. Identify all measuring instruments used in quality-affecting activities across your facility. Include sensors, transmitters, indicators, and any software that processes measurement data. This is your instrument register. The foundation of your calibration programme.
  2. Assign calibration intervals to each instrument based on the risk factors discussed in section 7: stability history, application severity, manufacturer guidance, and cost of error. Document the rationale for each interval decision.
  3. Set up a calibration register (spreadsheet, CMMS, or quality software), tracking instrument ID, description, location, calibration due date, certificate number, and OOT history. Configure automated alerts before due dates.
  4. Choose SAC-SINGLAS accredited labs for all calibrations in quality-affecting applications. Verify each lab's accreditation number and scope on sac.gov.sg before sending instruments. Confirm your specific parameters and ranges are within their accredited scope.
  5. Retain all calibration certificates as quality records. File them against the instrument ID in your register. ISO 9001 requires records to be retained for a defined period. Typically the instrument's lifetime plus a margin specified in your document control procedure.
  6. Train your team to respond correctly to an OOT finding: quarantine, impact assessment, root cause analysis, corrective action, and documentation. The OOT procedure should be a written procedure, not institutional knowledge held by one person.

The calibration system that follows these steps satisfies ISO 9001, IATF 16949, ISO 13485, GMP, and most sector-specific quality requirements without modification. The investment is in the structure and the discipline. The certificates from a SAC-SINGLAS accredited lab do the rest.

Frequently asked questions

What is calibration and why is it required in Singapore?

Calibration is the documented process of comparing a measuring instrument against a traceable reference standard to determine and record any deviation. It does not adjust the instrument, adjustment is a separate step. In Singapore, calibration is required by ISO 9001 (clause 7.1.5), IATF 16949, ISO 13485, GMP, and HACCP frameworks for all instruments used in quality-affecting activities. Without calibration, measurement data is untrustworthy and non-conformances during audits become the predictable consequence. Calibration is also required before an instrument's first use, after any impact or suspected damage, and at defined intervals throughout its operational life.

What is SAC-SINGLAS accreditation and why does it matter?

SAC-SINGLAS is Singapore's national laboratory accreditation scheme, operated by the Singapore Accreditation Council (SAC). Accreditation to ISO/IEC 17025 means the laboratory has been independently assessed on-site by SAC for technical competence. Traceability chain, uncertainty calculations, equipment, methods, and personnel. Accreditation number, scope, and validity are publicly verifiable at sac.gov.sg. SAC is a signatory to the ILAC Mutual Recognition Arrangement, so certificates from SAC-SINGLAS labs are accepted in the US, EU, Japan, Australia, and over 100 other economies. A certificate from a non-accredited source lacks this independent verification and its international recognition.

How often should instruments be calibrated?

The calibration interval is set by the instrument user, not the lab. ISO 9001 does not specify intervals. The decision should be risk-based: consider instrument stability and drift history, the severity of the application, the manufacturer's recommendation, and the cost and consequence of using an out-of-tolerance instrument. Typical practice in Singapore: 12 months for most industrial instruments; 6 months for critical safety applications; 3 months for precision pharmaceutical or semiconductor environments; up to 24 months for instruments with a long, documented history of stable calibration results. ISO 10012 provides a framework for interval adjustment using calibration history data.

What should a calibration certificate include?

A compliant calibration certificate under ISO/IEC 17025:2017 must include: the laboratory name and SAC-SINGLAS accreditation number; a unique certificate number and page count; the date of calibration and date of issue; instrument description with model and serial number; environmental conditions during calibration; the measurement method and reference standards used; calibration results at each test point; expanded measurement uncertainty with coverage factor k and confidence level; pass/fail against specification if requested; and the signature of an authorised person. A "CALIBRATED" sticker is an administrative label, not a quality record. It will not satisfy an ISO 9001, IATF, or GMP auditor asking for calibration evidence.

What do I do when an instrument fails calibration (out-of-tolerance)?

Follow the six-step OOT procedure: (1) Quarantine the instrument immediately (label it and remove from service. (2) Determine the affected period), the time since the last successful calibration. (3) Conduct an impact assessment. Review all measurements made during the affected period and assess whether any product or test result was compromised by the maximum possible error. (4) Perform root cause analysis (why did the instrument drift out of tolerance before the interval expired? (5) Take corrective action), repair, replace, or adjust and recalibrate. (6) Document everything. Impact assessment, root cause, corrective action, and outcome are all mandatory quality records under ISO 9001 clauses 8.7 and 10.2.

What is the difference between calibration, verification, and adjustment?

These three terms describe distinct actions that are often confused. Calibration is the documented measurement comparison against a traceable reference. It records what the instrument is actually doing relative to the true value. Verification is the act of comparing that calibration result against a defined tolerance or specification to determine whether the instrument is fit for its intended use. It answers the question "is this good enough?" Adjustment is the physical act of modifying the instrument to bring it closer to the true value. It is optional, performed only when calibration shows the deviation is unacceptable, and is always followed by re-calibration to document the new state.

Does Unitest provide SAC-SINGLAS accredited calibration for all instrument types?

Unitest Instruments holds SAC-SINGLAS accreditation no. LA-2023-0845-C covering electrical, temperature, pressure, dimensional, and flow parameters. All Unitest certificates state measurement uncertainty and carry full traceability to Singapore's National Metrology Centre (NMC). The full scope of accreditation is available to download from our accreditation page and is independently verifiable at sac.gov.sg. Contact us before sending instruments to confirm that your specific instrument type, parameter, and measurement range fall within our accredited scope. Scope-specific confirmation prevents delays and ensures the certificate you receive is fully accredited.

SAC-SINGLAS accredited laboratory mark
Written by Unitest Instruments

Unitest Instruments Pte. Ltd. is a SAC-SINGLAS accredited calibration laboratory (ISO/IEC 17025, no. LA-2023-0845-C) based in Singapore. We calibrate electrical, temperature, pressure, dimensional, and flow instruments for manufacturers, pharmaceutical companies, and regulated industries across Singapore and the region. All certificates carry full NMC traceability and stated measurement uncertainty. Ready for ISO 9001, GMP, and IATF 16949 audits.

SAC-SINGLAS accredited calibration for Singapore industry

Electrical, temperature, pressure, dimensional, and flow. NMC-traceable, ISO 9001 and audit-ready.

Verifiable at sac.gov.sg · LA-2023-0845-C